Metal artifact reduction in ultra
Scientific Reports volume 12, Article number: 15549 (2022) Cite this article
944 Accesses
1 Citations
1 Altmetric
Metrics details
Cone-beam computed tomography (CBCT) has been shown to be a powerful tool for 3D imaging of the appendicular skeleton, allowing for detailed visualization of bone microarchitecture. This study was designed to compare artifacts in the presence of osteosynthetic implants between CBCT and multidetector computed tomography (MDCT) in cadaveric wrist scans. A total of 32 scan protocols with varying tube potential and current were employed: both conventional CBCT and MDCT studies were included with tube voltage ranging from 60 to 140 kVp as well as additional MDCT protocols with dedicated spectral shaping via tin prefiltration. Irrespective of scanner type, all examinations were conducted in ultra-high-resolution (UHR) scan mode. For reconstruction of UHR-CBCT scans an additional iterative metal artifact reduction algorithm was employed, an image correction tool which cannot be used in combination with UHR-MDCT. To compare applied radiation doses between both scanners, the volume computed tomography dose index for a 16 cm phantom (CTDIvol) was evaluated. Images were assessed regarding subjective and objective image quality. Without automatic tube current modulation or tube potential control, radiation doses ranged between 1.3 mGy (with 70 kVp and 50.0 effective mAs) and 75.2 mGy (with 140 kVp and 383.0 effective mAs) in UHR-MDCT. Using the pulsed image acquisition method of the CBCT scanner, CTDIvol ranged between 2.3 mGy (with 60 kVp and 0.6 mean mAs per pulse) and 61.0 mGy (with 133 kVp and 2.5 mean mAs per pulse). In essence, all UHR-CBCT protocols employing a tube potential of 80 kVp or more were found to provide superior overall image quality and artifact reduction compared to UHR-MDCT (all p < .050). Interrater reliability of seven radiologists regarding image quality was substantial for tissue assessment and moderate for artifact assessment with Fleiss kappa of 0.652 (95% confidence interval 0.618–0.686; p < 0.001) and 0.570 (95% confidence interval 0.535–0.606; p < 0.001), respectively. Our results demonstrate that the UHR-CBCT scan mode of a twin robotic X-ray system facilitates excellent visualization of the appendicular skeleton in the presence of metal implants. Achievable image quality and artifact reduction are superior to dose-comparable UHR-MDCT and even MDCT protocols employing spectral shaping with tin prefiltration do not achieve the same level of artifact reduction in adjacent soft tissue.
In postoperative follow-up after joint arthroplasty, plain radiography is the primary imaging method due to ubiquitous availability, cost-efficiency and fast imaging results while pertaining relatively low radiation doses. For a more detailed analysis of suspected complications after surgery, additional CT may be necessary, albeit being associated with a higher dose penalty. However, in postoperative settings, artifacts caused by metal implants may hamper diagnostic accuracy for assessment of the implant itself, the implant-bone interface, as well as its adjacent tissue1,2. Typical metal artifacts include beam hardening and photon starvation: beam hardening occurs when polychromatic X-ray photons pass through dense objects resulting in stronger absorption of low energy photons, which causes hyperdense artifacts with adjacent dark streaks. In contrast, photon starvation artifacts manifest due to complete absorption of photons, which leads to hypodense streaks3,4,5. As a result, detecting complications in the presence of metal implants, such as secondary dislocation, areas of bone resorption or implant loosening as indicated by a surrounding radiolucent rim or even fluid collections in the soft tissue, can pose a significant challenge.
Different approaches for metal artifact reduction (MAR) have been evaluated predominantly for conventional gantry-based, multidetector CT (MDCT) scanners in the past6,7. Photon starvation can be diminished by increasing the tube current in order to augment the number of photons in the X-ray beam. Increased tube voltage and thus higher photon energy leads to a higher penetration rate of dense material. At higher tube voltages, image noise and photon starvation can be minimized at the cost of decreased tissue contrast. However, reduction of metal artifacts at the expense of higher radiation doses is debatable, especially in young patients and patients with recurring examinations8. A similar effect can be seen when employing tin prefiltration, which increases photon penetration by reducing the amount of low-energy photons and therefore hardening the X-ray-beam7,9. While these protocol-based MAR approaches need to be established in advance of image acquisition, algorithms such as iterative reconstruction techniques can be executed retrospectively without negatively effecting the radiation dose. On the downside, iterative reconstruction algorithms may introduce secondary artifacts and have been reported to alter image information in general10,11. Furthermore, image data may be lost near the metal edge by interpolation12. Apart from optimization of metal implants and scan protocols, other approaches to reduce suchlike artifacts include model-based data correction, and image-based postprocessing13.
Due to the excellent image quality of bone tissue at relatively low dose, cone-beam CT (CBCT) has more recently developed a growing niche in musculoskeletal imaging. In trauma assessment, CBCT has already emerged as a viable alternative for depiction of the upper and lower extremity14,15,16,17,18. In a post-treatment setting, however, the value of non-dental CBCT systems has not been thoroughly investigated thus far. Especially in patients with limited mobility after osteosynthesis, the prone position and arm elevation required for optimized imaging conditions in wrist scans can be problematic with gantry-based MDCT scanners. Addressing this limitation with the investigated gantry-free multi-purpose scanner's option for tableside positioning of the upper extremity can be advantageous with regards to image quality and radiation dose19.
The goal of this study was to investigate the metal artifact reduction capabilities of a twin robotic X-ray system with an ultra-high-resolution (UHR)-CBCT scan mode in an experimental setting. Therefore, we compared the scanner's performance to UHR-MDCT in cadaveric wrist scans after arthroplasty, aiming to find the best compromise between radiation dose and image quality.
Our study was approved by the Institutional Review Board of the University of Würzburg, Germany. During their lifetime, donors volunteered their body to the university's anatomical institute for study and research purposes. Additional written informed consent was not required.
All methods were performed in accordance with the relevant guidelines and regulations. The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.
Two trauma surgeons performed volar locking plate fixation (Aptus, Medartis, Basel, Switzerland) on a formalin-fixed cadaveric specimen's distal radius. The wrist was then examined employing a multipurpose, twin robotic X-ray system for CBCT imaging and a high-end MDCT scanner. MDCT studies were performed in the so-called "superman stance" with the specimen in prone position and elevation of the respective arm above the head. CBCT studies were conducted in supine position with the upper extremity abducted at a 90° angle for a tableside scan position.
The employed multi-use X-ray system with implemented 3D CBCT mode (Multitom Rax, Siemens Healthineers, Erlangen, Germany) is equipped with two telescopic arms carrying the X-ray tube and flat-panel detector and is connected to ceiling rails with three translational and two rotational degrees of freedom. The X-ray tube is capable of providing currents from 0.5 to 800 mAs and voltages from 40 to 150 kV. The input field of the flat panel detector measures 23 cm × 23 cm with a 3D matrix of 1440 × 1440 un-binned pixels, which results in an effective pixel size of 148 μm. At the same time, the input field limits the CBCT's maximum field of view since table movement and helical acquisition are not feasible with this system. For the acquisition of 3D projection data in CBCT scan mode, the arms move simultaneously and in synchronized fashion along predefined scanning trajectories. The tableside trajectory for upper extremity imaging has a sweep angle of 200° around the isocentre with an asymmetric source-to-image distance of 115 cm (tube-to-isocentre distance 85 cm, detector-to-isocentre distance 30 cm). For the purpose of the study, the wrist was additionally scanned with a high-end gantry-based MDCT scanner and multiple protocols in single-energy mode (Somatom Force, Siemens Healthineers).
Different scan protocols were evaluated while varying prefiltration, tube voltage and tube current-exposure time products. Thirty-two scans of the cadaveric wrist with volar locking plate fixation were performed in UHR mode. A total of 15 UHR-CBCT scans were evaluated with tube voltage ranging from 60 to 133 kVp and tube current-exposure-time products of 0.6 mAs, 1.2 mAs and 2.5 mAs. Of note, the CBCT system's standard hardware does not include the option for tin prefiltration. The protocols for matching UHR-MDCT scans (15 matching MDCT protocols) were set up with tube voltage ranging from 70 to 140 kVp with reference mAs designed to equal CTDI values of their CBCT counterparts. Additionally, two protocols (100 kVp and 150 kVp) with 0.4 mm tin prefiltration were employed, resulting in a total of 19 MDCT protocols. Detailed scan parameters for CBCT and MDCT protocols are summarized in Table 1. For comparison of applied radiation doses between CBCT and MDCT, the volume computed tomography dose index for a 16 cm phantom (CTDIvol) was evaluated. Dose-length products and CTDIvol were noted to estimate the radiation dose in MDCT. CTDIvol-equivalent values for CBCT were computed by multiplying the dose-area product by a linear scaling factor provided by the manufacturer based on previous phantom measurements. The scaling factor was determined in advance for every combination of acquisition geometry, tube voltage and beam filtration (e.g. copper filter). The dose-length product was measured in five chambers of a conventional polymethyl methacrylate dosimetry phantom (IEC 60601-2-44:2009) with a diameter of 160 mm, a length of 300 mm and a standard dosimetry system (Nomex Dosimeter, PTW, Freiburg im Breisgau, Germany) with a 300 mm ionisation chamber. Standard weighting scheme for dose measurements was applied to acquire volume dose-length product values (DLPvol). CTDIvol values were calculated by dividing DLPvol values by the field of view in the z-direction, which is equivalent to the beam width. Finally, the scaling factor was computed by dividing CTDIvol by dose-area product values. Dose-area products for all examinations were extracted from the automatically created scan report.
Scanner-side raw data reconstruction was performed employing a dedicated high-resolution kernel for very sharp depiction of osseous structures (Ur77; Siemens Healthineers) following the clinical standard for post-processing of MDCT examinations at our institution. Multiplanar reconstructions were carried out for CBCT and MDCT using dedicated 3D processing software (syngo.via View&GO and syngo.via, both Siemens Healthineers). For reconstruction of UHR-CBCT scans additional iterative metal artifact reduction (MAR) was employed. Reconstruction parameters for axial, coronal and sagittal planes were set with slice thickness of 1.0 mm, increment of 0.5 mm, field of view of 80 mm, and image matrix of 1024 × 1024 pixels irrespective of dose protocol and scanner. Window width and level of 3000 and 1000 Hounsfield units (HU) were predefined for optimal bone depiction. However, readers were allowed to alter window settings as personally required.
Images were evaluated independently in randomized and blinded fashion using clinical picture archiving and communication software (Merlin, Phoenix-PACS, Freiburg im Breisgau, Germany) by seven radiologists with varying levels of experience in musculoskeletal imaging (ASK 8, TSP 2, KSL 7, VH 6, RH 4, HH 6 and JPG 5 years of experience). No time restriction was imposed for reading. The readers were asked to rate the image quality with regards to osseous and soft tissue as well as the extent of artifacts using a seven-point scale (7 = excellent/no artifacts, 6 = very good/near absence of artifacts, 5 = good/mild artifacts, 4 = satisfactory/moderate artifacts, 3 = fair, /considerable artifacts 2 = poor/severe artifacts, 1 = very poor/not diagnostic due to artifacts). Based on these results we calculated figure-of-merit (FOM) values for bone and artifact ratings in order to characterize scan protocol performance and to rank image quality relative to radiation dose using the following formula:
Objective image analysis was performed by a radiologist with 5 years of experience in musculoskeletal imaging (JPG). Circular regions-of-interest (ROI) were manually positioned in surrounding air and soft tissue as reference measuring signal attenuation in mean HU. Artifacts were quantified by placing a ROI in the most pronounced hypo- and hyperdense artifact areas as well as in the artifact-impaired soft tissue. Artifact measurements were performed three times, respectively, and averaged to guarantee high measurement accuracy and data consistency.
Dedicated software was used for descriptive statistics and data analyses (SPSS Statistics Version 28, IBM, Armonk, New York, USA). P values less than 0.05 were considered to indicate statistical significance. Kolmogorov–Smirnov tests were applied to analyze continuous data for normal distribution. If normally distributed, continuous data is presented as mean ± standard deviation. Means of normally-distributed variables were compared with one-way ANOVA and Dunnett-T3 post-hoc tests with p values adjusted for multiple comparisons. Mean ranks of categorical items were compared with Friedman tests and post-hoc analysis of homogenous subsets. Fleiss kappa was calculated to investigate the interrater reliability in subjective image quality assessment. Agreement was interpreted according to Landis and Koch (1.00–0.81 = almost perfect; 0.80–0.61 = substantial; 0.60–0.41 = moderate; 0.40–0.21 = fair; 0.20–0.00 slight; < 0.00 poor agreement)20.
Thirty-two scan protocols with varying prefiltration, tube voltage and current–time product were applied in this study (17 UHR-MDCT, 15 UHR-CBCT). Without automatic tube current modulation or tube potential control, the highest and lowest radiation doses achieved in conventional MDCT were 75.2 mGy (with 140 kVp and 383.0 effective mAs) and 1.3 mGy (with 70 kVp and 50.0 effective mAs). Using the pulsed image acquisition method of the CBCT scanner, CTDIvol of ultra-low-dose scans was 2.3 mGy (with 60 kVp and 0.6 mean mAs per pulse), while the highest possible dose was 61.0 mGy (with 133 kVp and 2.5 mean mAs per pulse). CTDIvol values are summarized in Table 1.
For visual comparison of image quality and metal artifact intensity with various combinations of tube potential and current on both scanners, representative images of the distal forearm after palmar plate osteoplasty are presented in Fig. 1 (UHR-MDCT) and Fig. 2 (UHR-CBCT). ROI-based quantification of signal attenuation in hyperdense beam hardening (all p < 0.036) and hypodense photon starvation artifacts (all p < 0.008) yielded favorable results for UHR-CBCT protocols with at least 80 kVp compared to UHR-MDCT imaging (Fig. 3). While providing inferior results in comparison with UHR-CBCT protocols, tin prefiltration lowered the extent of hyperdense and hypodense artifacts considerably compared to standard UHR-MDCT (all p < 0.003). With tin prefiltration activated, 150 kVp scans displayed stronger artifact reduction than 100 kVp imaging (p = 0.034). Impairment of adjacent soft tissue by artifacts was also reduced significantly with tin-prefiltered scan protocols (all p < 0.029), albeit UHR-CBCT with 80 kVp or more allowed for the best depiction of artifact-impaired soft tissue in this study (all p < 0.001). No significant differences were ascertained between UHR-CBCT scan protocols with at least 80 kVp (all p > 0.366). Results of objective artifact intensity assessment for both scanners are summarized in Table 2. In addition, boxplot diagrams are provided to illustrate the differences between UHR-MDCT and UHR-CBCT for each type of artifact (Fig. 4).
Bone image quality and intensity of metal artifacts by palmar plate osteosynthesis in conventional ultra-high-resolution MDCT scan protocols.
Bone image quality and intensity of metal artifacts by palmar plate osteosynthesis in ultra-high-resolution CBCT scan protocols.
Region of interest (ROI) placement for objective assessment of metal artifact intensity. Note—A = hyperdense artifacts; B = hypodense artifacts; C = artifact-impaired soft tissue.
Boxplots with signal attenuation in artifacts and artifact-impaired soft tissue. Note—Boxplots (median and 50% of cases within the boxes) illustrate the corrected signal attenuation in Hounsfield units (HU) for hyperdense artifacts, hypodense artifacts and artifact-impaired soft tissue in cone-beam computed tomography (CBCT) and multidetector computed tomography (MDCT) with different levels of tube voltage (kVp = kilovoltage peak).
FOM analysis (Table 3) showed that UHR-CBCT protocols performed better than the dose-comparable UHR-MDCT protocols, owing primarily to relatively good ratings for protocols with low radiation exposure (except for a portion of 60 kVp scans, where image quality in bone tissue was considered insufficient for clinical use). The best protocol overall was the current clinical standard (80 kVp, 0.6 mean mAs per pulse). Despite favorable ratings, the higher-voltage protocols on both scanners achieved poorer FOM values due to their increase in radiation dose. Of note, tin-filtered UHR-MDCT scans received considerably better FOM values than standard UHR-MDCT examinations for bone image quality and artifact intensity. Homogeneous subset analysis of scan protocols based on tube voltage groups showed higher mean rank values for UHR-CBCT studies with at least 80 kVp regarding bone tissue delineation (Table 4) and artifacts (Table 5) (all p < 0.050). Interrater reliability of seven radiologists was substantial for bone and moderate for artifact assessment, as was indicated by Fleiss kappa of 0.652 (95% confidence interval 0.618–0.686; p < 0.001) and 0.570 (95% confidence interval 0.535–0.606; p < 0.001), respectively.
In this experimental multi-observer study, we evaluated the potential of UHR-CBCT imaging in the setting of a present metal implant in comparison to a third-generation dual-source MDCT scanner with UHR option. To this effect, a total of 32 protocols with varying acquisition parameters were applied on a human cadaveric specimen. For tube potentials of 80 kVp and more, we were able to show that UHR-CBCT displayed superior performance regarding image artifacts as per objective analysis, and regarding image quality as per subjective analysis by seven radiologists. Furthermore, applied radiation dose was favorable in UHR-CBCT compared to UHR-MDCT protocols with similar image quality in bone and soft tissue.
The employed protocols for MDCT consisted of conventional CT imaging with tube voltage ranging from 70 to 140 kVp, as well as two protocols employing tin prefiltration for dedicated metal artefact reduction (Sn 100 kVp and Sn 150 kVp). In order to maintain comparability between CBCT and MDCT, all scans were performed in UHR mode and without employing dual-energy protocols. In other settings, spectral shaping via tin prefiltration has been shown to reduce applied dose considerably, especially in the case of obese patients when compared to dual-energy imaging and employment of virtual monochromatic reconstructions21. Notably and in contrast to UHR-CBCT, with current third-generation dual-source CT scanners, the necessity to employ an additional comb filter for UHR imaging effectively forces the choice between employing UHR acquisition and additional iterative MAR. While UHR-MDCT protocols employing spectral shaping via tin prefiltration significantly reduced metal artifacts compared to standard UHR-MDCT, UHR-CBCT with 80 kVp or more did allow for even superior artifact reduction and assessability of adjacent soft tissue. Also, regarding subjective image quality, UHR-CBCT protocols with tube voltage of 80 kVp or more displayed superior performance as compared to UHR-MDCT imaging. One explanation for the CBCT's performance lies in its inherent superior dose efficiency which is realized by the system's unique acquisition geometry: While regular gantry-based scanners operate with symmetric intervals between the X-ray tube and patient, as well as patient and detector, the employed CBCT's two telescopic arms enable an asymmetric source-to-image distance with low magnification that improves spatial resolution by counteracting the limiting effect of the focal spot size22.
In light of recent publications demonstrating the potential and advantages of tableside UHR-CBCT imaging of the appendicular skeleton in trauma settings19,22, the reported findings are noteworthy. With regards to applied average dose, a recent meta-analysis by Nardi et al. states that CBCT promises a significant reduction over MDCT23. However, CBCT image quality is dependent on placement of the object of interest in the system's iso-center due to the limited number of acquired image frames. This fact is more pronounced than in MDCT, potentially necessitating dedicated training and demanding explicit care by the radiographers.
CT imaging with metal implants present has been part of clinical routine for decades. However, artifact reduction in adjacent tissue is oftentimes of utmost importance, as even subtle findings such as fissure fractures require specific follow-up treatment up to secondary surgical revision. While high image quality is essential in these scans, dose reduction efforts can have a contrary effect on image quality. As conventional UHR-MDCT protocols were found to be clearly inferior and even UHR-MDCT protocols employing spectral shaping were not on par with UHR-CBCT imaging, the authors suggest future studies in clinical settings to further evaluate this promising imaging technique. In addition, dedicated studies quantifying the effect of the iterative metal artifact reduction algorithm with CBCT are mandated.
Some limitations ought to be mentioned regarding this study. For one, scans were limited to a single cadaveric specimen and to one type of implant. Thus, the impact of implant composition and size on resulting artifacts was not evaluated. Even though patients may well tolerate the required acquisition time of 14 s in CBCT due to comfortable tableside positioning, possible motion artifacts may be a limiting factor for clinical routine23. It lies in the nature of the study design, however, that motion artifacts did not play a role for this investigation. Furthermore, formalin fixation has been reported to invoke demineralization of the bone over time, impeding image quality irrespective of scanner and scan protocol24,25. Even though observers were blinded to the type of scan (UHR-CBCT vs. UHR-MDCT), a certain level of bias is conceivable due to the typical image impression of each imaging modality. Lastly, Multitom Rax (Siemens Healthineers) is not commercially available in all countries, therefore reproducibility of results may be limited.
With the cone-beam CT scan mode of a gantry-free twin robotic X-ray system, ultra-high-resolution imaging of the appendicular skeleton in the presence of metal implants can be realized. Whilst the compared third-generation dual-source multidetector CT system does not support iterative metal artifact reduction algorithms in UHR-mode, achievable image quality and artifact reduction by means of cone-beam CT are superior to dose-comparable conventional UHR-multidetector CT and even multidetector CT protocols employing spectral shaping with tin prefiltration.
The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.
Cone-beam computed tomography
Volume computed tomography dose index for a 16 cm phantom (mGy)
Metal artifact reduction
Multidetector computed tomography
Region of interest
Ultra-high-resolution
Wellenberg, R. H. H. et al. Metal artifact reduction techniques in musculoskeletal CT-imaging. Eur. J. Radiol. 107, 60–69 (2018).
Article CAS Google Scholar
Chandrasekar, S. et al. Combined dual-energy and single-energy metal artifact reduction techniques versus single-energy techniques alone for lesion detection near an arthroplasty. AJR Am. J. Roentgenol. 215, 425–432 (2020).
Article Google Scholar
Bushberg, J. T. The AAPM/RSNA physics tutorial for residents X-ray interactions. Radiographics 18, 457–468 (1998).
Article CAS Google Scholar
GroßeHokamp, N. et al. Artifact reduction from dental implants using virtual monoenergetic reconstructions from novel spectral detector CT. Eur. J. Radiol. 104, 136–142 (2018).
Article Google Scholar
Barrett, J. F. & Keat, N. Artifacts in CT: recognition and avoidance. Radiographics 24, 1679–1691 (2004).
Article Google Scholar
Boas, F. E. & Fleischmann, D. CT artifacts: Causes and reduction techniques. Imaging Med 4, 229–240 (2012).
Article Google Scholar
Hackenbroch, C. et al. Dose reduction in dental CT: A phantom study with special focus on tin filter technique. AJR Am. J. Roentgenol. 215, 945–953 (2020).
Article Google Scholar
Gupta, A., Subhas, N., Primak, A. N., Nittka, M. & Liu, K. Metal artifact reduction: Standard and advanced magnetic resonance and computed tomography techniques. Radiol. Clin. North Am. 53, 531–547 (2015).
Article Google Scholar
Zhou, W. et al. Reduction of metal artifacts and improvement in dose efficiency using photon-counting detector computed tomography and tin filtration. Invest. Radiol. 54, 204–211 (2019).
Article CAS Google Scholar
Chang, W. et al. Assessment of a model-based, iterative reconstruction algorithm (MBIR) regarding image quality and dose reduction in liver computed tomography. Invest. Radiol. 48, 598–606 (2013).
Article CAS Google Scholar
Messerli, M. et al. Impact of advanced modeled iterative reconstruction on coronary artery calcium quantification. Acad. Radiol. 23, 1506–1512 (2016).
Article Google Scholar
Meyer, E., Raupach, R., Lell, M., Schmidt, B. & Kachelrieß, M. Normalized metal artifact reduction (NMAR) in computed tomography. Med. Phys. 37, 5482–5493 (2010).
Article Google Scholar
Gjesteby, L. et al. Metal artifact reduction in CT: Where are we after four decades?. IEEE Access 4, 5826–5849 (2016).
Article Google Scholar
Luetkens, K. S. et al. Dose reduction potential in cone-beam CT imaging of upper extremity joints with a twin robotic X-ray system. Sci. Rep. 11, 20176 (2021).
Article ADS CAS Google Scholar
Carrino, J. A. et al. Dedicated cone-beam CT system for extremity imaging. Radiology 270, 816–824 (2014).
Article Google Scholar
Demehri, S. et al. Assessment of image quality in soft tissue and bone visualization tasks for a dedicated extremity cone-beam CT system. Eur. Radiol. 25, 1742–1751 (2015).
Article CAS Google Scholar
Neubauer, J. et al. Comparison of the diagnostic accuracy of cone beam computed tomography and radiography for scaphoid fractures. Sci. Rep. 8, 3906 (2018).
Article ADS Google Scholar
Yang, T. W. et al. Diagnostic performance of cone-beam computed tomography for scaphoid fractures: A systematic review and diagnostic meta-analysis. Sci. Rep. 11, 2587 (2021).
Article CAS Google Scholar
Grunz, J. P. et al. Twin robotic X-ray system in small bone and joint trauma: impact of cone-beam computed tomography on treatment decisions. Eur. Radiol. 31, 3600–3609 (2021).
Article Google Scholar
Landis, J. R. & Koch, G. G. The measurement of observer agreement for categorical data. Biometrics 33, 159 (1977).
Article CAS Google Scholar
Gordic, S. et al. Ultralow-dose chest computed tomography for pulmonary nodule detection: First performance evaluation of single energy scanning with spectral shaping. Invest. Radiol. 49, 465–473 (2014).
Article Google Scholar
Grunz, J. P. et al. 3D cone-beam CT with a twin robotic X-ray system in elbow imaging: Comparison of image quality to high-resolution multidetector CT. Eur. Radiol. Exp. 4, 52 (2020).
Article Google Scholar
Nardi, C. et al. Motion artefacts in cone beam CT: an in vitro study about the effects on the images. Br. J. Radiol. 89, 20150687 (2016).
Article Google Scholar
Fonseca, A. A., Cherubini, K., Veeck, E. B., Ladeira, R. S. & Carapeto, L. P. Effect of 10% formalin on radiographic optical density of bone specimens. Dentomaxillofac Radiol. 37, 137–141 (2008).
Article CAS Google Scholar
Burkhart, K. J. et al. Influence of formalin fixation on the biomechanical properties of human diaphyseal bone. Biomed. Tech. (Berl) 55, 361–365 (2010).
Article Google Scholar
Download references
We wish to thank the individuals who donated their body for research and educational purposes.
Open Access funding enabled and organized by Projekt DEAL.
These authors contributed equally: Andreas Steven Kunz and Theresa Sophie Patzer.
Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080, Würzburg, Germany
Andreas Steven Kunz, Theresa Sophie Patzer, Jan-Peter Grunz, Karsten Sebastian Luetkens, Viktor Hartung, Robin Hendel, Thorsten Alexander Bley & Henner Huflage
Department of Trauma, Hand, Plastic and Reconstructive Surgery, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080, Würzburg, Germany
Tabea Fieber
Orthopedic Clinic König-Ludwig-Haus, Julius-Maximilians-Universität Würzburg, Brettreichstr. 11, 97070, Würzburg, Germany
Franca Genest
Institute of Anatomy and Cell Biology, University of Würzburg, Koellikerstr. 6, 97070, Würzburg, Germany
Süleyman Ergün
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
A.S.K. and T.S.P. contributed equally as first authors, analyzing all data and preparing the manuscript. Observer analysis was performed by A.S.K., T.S.P., K.S.L., V.H., R.H., H.H., and J.P.G. T.F. and F.G. supported preparation of the manuscript and figures. J.P.G. performed statistical analysis and revised the manuscript. T.A.B. and S.E. contributed to preparation of the manuscript and provided quality control. S.E. prepared the cadaveric specimen, while T.F. and F.G. performed palmar plate osteoplasty. H.H. designed and supervised the study. All authors read and approved the final manuscript.
Correspondence to Andreas Steven Kunz.
J.P.G was funded by the Interdisciplinary Center of Clinical Research Würzburg, Germany [grant number Z-2/CSP-06] and serves as a research consultant for Siemens Healthineers. The Department of Diagnostic and Interventional Radiology receives research funding by Siemens Healthineers. The authors of this manuscript declare no further relationships with any companies, whose products or services may be related to the subject matter of the article.
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
Reprints and Permissions
Kunz, A.S., Patzer, T.S., Grunz, JP. et al. Metal artifact reduction in ultra-high-resolution cone-beam CT imaging with a twin robotic X-ray system. Sci Rep 12, 15549 (2022). https://doi.org/10.1038/s41598-022-19978-9
Download citation
Received: 05 March 2022
Accepted: 07 September 2022
Published: 16 September 2022
DOI: https://doi.org/10.1038/s41598-022-19978-9
Anyone you share the following link with will be able to read this content:
Sorry, a shareable link is not currently available for this article.
Provided by the Springer Nature SharedIt content-sharing initiative
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.